In the last tutorial on CRO I gave you a detailed idea about the various components of an Oscilloscope concerned with the positioning and magnification of an input signal and the role of probes was also explained. Now since you have been acquainted with the primary knobs of a Cathode Ray Oscilloscope it’s time to dive deeper and to know more about the other knobs which are helpful in further observation of a signal.
On/Off: This is required to turn the CRO on or off as per the requirement.
Fig. 1: Image showing On/Off Button of CRO
Channel Select: As we know most of the CRO can work simultaneously with two signals and thus there is a toggle switch present in the controls regarding the selection, i.e., if both the channels are used or only one is used. So, selection must be done accordingly. If both the channels are used the selection should be ‘Dual’ and if only one channel is used it should be either ‘channel 1’ or ‘channel 2’ depending on the connection.
Fig. 2: Image showing Channel Select Buttons of CRO
Trigger Hold-Off: The trace in a CRO is performed by the sweeping of electrons for a small period of time on the phosphorous screen by the voltage generated due to the signal. This voltage is known as sweep voltage. When the ‘sweep’ is completed the voltage returns to its original value and thus forcing the beam of electrons to move back. This process is called ‘retrace’, when the retrace is completed again sweep starts. Hold-Off determines the time that tends to pass between ‘retrace’ and ‘sweep’. Initially it is set to zero, but to avoid the certain peaks and noises in the signal the sweep time can be increased to obtain a stable signal at the display.
Fig. 3: Image showing Trigger Hold-Off Button of CRO
AC/DC/GD: This is a toggle switch to tell the CRO about the filtering of the i/p signal channel. This is present one for each channel.
· AC (Alternating Current Coupling): If the switch is at AC then the signal will go through a RC-filter to remove its DC components.
·DC (Direct Current Coupling): If the switch is at DC position then the signal will not go through a RC-filter and the DC components will persist in the signal.
·GD (Ground): If the switch is at the GD position then the input signal is shorted and thus a horizontal line is seen in the display symbolizing no potential difference between ground and input signal.
Focus: This rotating knob is used to improve the sharpness of an image of the signal on the screen by adjusting the focus of the electronic beam. Thus, it helps in taking the proper measurements and to study the correct shape of the signal by removing the blurriness of the signal.
Intensity: This knob is used to increase the intensity of the input signal for better examination of the displayed output on the screen. There is a common misconception regarding this knob to be a brightness control feature but it is not so, this increases the number of electrons in the beam so that more and more electrons collide the phosphorus screen for better view of the applied input signal. This knob is marked as ‘Intens’ on the CRO.
Fig. 4: Image showing Intensity and Focus Buttons of CRO
+/- : This determines if the signal is triggered during the positive flank or the negative flank. Now, the many of you must be wondering ‘what is triggering?’, it is the synchronization of the horizontal sweep of the CRO with the proper point of the signal. If the button is on descending slope of the signal is used while triggering otherwise the ascending slope of the signal is used.
At/Norm.: Using this button you can select between Automatic triggering (At.) and manual triggering level selection.
Level: When the triggering is set to manual mode this wheel is used to set the level of manual triggering, the +/- button will make sure ascending level triggering or descending level triggering is used.
Ext.: This button is used when the timing of the CRO is not determined by the input waveform but from some another external waveform. There is an input connector beside the button where the external input is connected for the display of the waveform.
Fig. 5: Image showing Ext Button of CRO
Other Functions and Sampling Example
Trig.: This button has four options to select and is used to filter the triggering signal of the Oscilloscope.
{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C} · AC: The DC component of the trigger signal is filtered out i.e. the zero frequency signal have no effect on the triggering signal.
{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C} · {C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}DC: The DC selection will let all the signal starting at zero frequency initiate or influence the triggering of the input signal.
{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C} · {C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}LF: The high frequency signals are filtered out to remove the unwanted noise or spikes in the triggering.
{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C} ·HF: The Low frequency signals components are filtered out.
{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C}{C} · ~ : This position of the switch will let the triggering of the input signal by the power supply frequency of the Oscilloscope.
Fig. 6: Image showing Buttons of CRO for Other Functions
Invert: This button inverts the polarity of the signal input thus the positive part of the signal becomes negative and the negative part of the signal becomes positive. Thus there is an inversion on the display of the Oscilloscope. This button is present for each of the channel separately.
Thus we have gone through all most all the knobs present in an Oscilloscope which is required for the observational analysis of a signal. Now, let us go through a simple example to learn the practical application of these buttons and knobs.
Example
Fig. 7: Image showing Buffer, Sampler and Filter Circuit Board
For instance, we are doing the natural sampling and reconstruction of a sinusoidal signal using the DCL 01 Falcon kit. This kit is basically used for the sampling of analog signals in various ways and their reconstruction. It basically consists of functional block namely Function generator, Sampling control, Logic, Clock section, Sampling circuitry and Filter section.
Fig. 8: Block Diagram of Buffer, Sampler and Filter
This is the block diagram of the circuit involved in the experiment and the connections are done according to it. Now when the connections are done (which is very easy because of the presence of each component in blocks in the kit), we intend to see the wave forms at different stages of the experiment. We will connect one of the connectors of one side of the probe with the function generator at ‘point 1’ and other connector of the same side to the ground. The other side is connected to the Oscilloscope to either of the channels. We will be using the vertical position and voltage scale to bring changes in the positioning and magnification of the signal. The display obtained is shown in figure below.
Fig. 9: Signal Diagram showing Sampling Frequency
Now we want to check the waveform after the sampling of the signal. So we connect the CRO probe with the ‘point 2’ of the circuit and see the waveform generated in the display screen. Then the waveform generated is shown below. We can magnify the signal by decreasing the voltage scale and also could change the position of the signal on the display by the help of vertical positioning knob.
Fig. 10: Signal Diagram showing Output Wave of Natural Sampling
Now for the reconstruction of the signal the sampled signal is passed through 2nd order lowpass Butterworth filter. The waveform obtained is similar to input sinusoidal signal but far to perfection, also there is flickering in the display which can be controlled by the hold off knob. The waveform obtained is shown below. The waveform can be improved by further passing it through 4th order Butterworth lowpass filter.
Fig. 11: Signal Diagram showing Reconstructed Wave of Natural Sampling